Operation – KROHNE OPTISONIC 7300 EN User Manual
Page 87
OPERATION
6
87
OPTISONIC 7300
www.krohne.com
09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en
C
simulation VoS
Sensor electronics simulates
measurement of velocity of sound.
C
simulation
fieldbus
Simulation is active on the fieldbus
values.
I
counter 1 (or 2,
3) stopped
Counter has stopped.
Reset counter in menu C5.9.1
(or C5.9.2, C5.9.3).
I
power fail
The device was not in operation for
an unknown period of time.
Temporary power failure, counters
did not run during the failure.
I
control input A
(or B) active
Information only.
No action needed.
I
over range
display 1 (or 2)
1
st
row on 1
st
(or 2
nd
) measurement
page is limited by parameter setting.
Extend upper or lower limit for
limitation in menu C5.3.4 (or C5.4.4).
I
backplane
sensor
Incompatible data sensor on
backplane.
I
backplane
settings
Incompatible data on backplane.
I
backplane
difference
Different data on backplane and
display.
I
optical
interface
GDC IR interface is operational, local
display can not be used.
The keys are ready for operation
again 60 seconds after the end of
data transfer / removal of GDC IR
interface.
I
write cycles
overfl.
The maximum number of write
cycles of the EEPROM or FRAMS on
the Profibus PCB has been
exceeded.
I
baudrate
search
Searching for the baudrate of the
Profibus DP interface.
I
no data
exchange
No data exchange between converter
and Profibus.
I
start up
The converter is starting up and wait
time is required.
Error
Error
Error
Error
code
code
code
code
Error
Error
Error
Error
message
message
message
message
Description
Description
Description
Description
Error handling
Error handling
Error handling
Error handling