OPERATION
6
65
OPTISONIC 7300
www.krohne.com
09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en
6.3 Function tables
6.3.1 Menu A, quick setup
INFORMATION!
•
The following tables describe the functions of the standard device with HART
®
connection.
The functions for Modbus, Foundation Fieldbus and Profibus are described in detail in the
corresponding supplementary instructions.
•
Depending on the device version, not all functions are available.
No.
No.
No.
No.
Function
Function
Function
Function
Settings / descriptions
Settings / descriptions
Settings / descriptions
Settings / descriptions
A1
language
Select: English / Français / Deutsch
A2
tag
Measuring point identifier (Tag no.) (also for HART
®
operation),
appears in the LCD display header (up to 8 digits).
A3
reset
A3.1
reset errors
Reset? Select: no/yes
A3.2
counter 1
Reset counter? Select: no / yes (available if activated in C5.9.1)
A3.3
counter 2
Reset counter? Select: no / yes (available if activated in C5.9.2)
A3.4
counter 3
Reset counter? Select: no / yes (available if activated in C5.9.3)
A4
GDC IR interface
This function requires an optical GDC adapter connected to the LCD
display. After 60 seconds without a connection being established or
after the adapter is removed, the function is exited and the optical
keys are active again.
Select: break (exit function without connection)
Activate (the IR interface adapter and interrupt the optical keys)