DESCRIPTION OF HART INTERFACE
9
113
OPTISONIC 7300
www.krohne.com
09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en
9.4 Inputs/outputs and HART
®
dynamic variables and device variables
The signal converter is available with various in-/output combinations.
The connection of the terminals A…D to the HART
®
dynamic variables PV, SV, TV and QV
depends on the device version.
PV = Primary Variable; SV = Secondary Variable; TV = Third Variable; QV = Quarternary Variable
The signal converter can provide up to 14 measurement-related values. These values are
accessible as so-called HART
®
device variables and can be connected to the HART
®
dynamic
variables. The availability of these variables depends on the device versions and the settings.
Code = device variable code
Device variables
For the dynamic variables connected to the linear analogue outputs (for current and/or
frequency) are assigned by selecting the measurement for the related outputs. Only linear
device variables can be assigned in this case.
For dynamic variables not connected to linear analogue outputs both linear and totaliser device
variables can be assigned.
Signal converter version
HART
®
dynamic variable
PV
SV
TV
QV
Basic I/O, connection terminals
A
D
-
-
Modular I/O and Ex i I/O, connection terminals
C
D
A
B
HART
®
device variable
Code Type
Explanations
volume flow
20
linear
corrected volume flow
21
linear
mass flow
22
linear
molar mass
23
linear
flow speed
25
linear
velocity of sound
26
linear
signal gain
27
linear
diagnosis 1
28
linear
Function and availability depends on diagnosis 1 value
setting.
diagnosis 2
29
linear
Function and availability depends on diagnosis 2 value
setting.
diagnosis 3
30
linear
Function and availability depends on diagnosis 3 value
setting.
counter 1 (C)
6
totaliser
Valid for Basic I/O option only.
counter 1 (B)
13
totaliser
Valid for Modular I/O and Ex i I/O options only.
counter 2 (D)
14
totaliser
-
counter 3 (A)
12
totaliser
Valid for Modular I/O and Ex i I/O options only.