4

ELECTRICAL CONNECTIONS

48

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Current input active, modular I/Os

• U

int, nom

= 24 VDC

• I ≤ 22 mA
• I

max

≤ 26 mA (electronically limited)

• U

0, min

= 19 V at I ≤ 22 mA

• no

no

no

no HART®

• X designates the connection terminals A or B, depending on the version of the signal

converter.

Current input passive, modular I/Os

• U

ext

≤ 32 VDC

• I ≤ 22 mA
• I

max

≤ 26 mA

• U

0, max

= 5 V at I ≤ 22 mA

• X designates the connection terminals A or B, depending on the version of the signal

converter.

Figure 4-22: Current input active IIn

a

1 Signal
2 2-wire transmitter (e.g. temperature)

Figure 4-23: Current input passive IIn

p

1 Signal
2 2-wire transmitter (e.g. temperature)