DESCRIPTION OF HART INTERFACE

9

111

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

9.3.2 Multi-Drop connection (2-wire connection)

In the case of a multi-drop connection, up to 15 devices may be installed in parallel (this signal
converter and other HART

®

devices).

The current outputs of the devices must be passive!

1 Primary Master
2 HART

®

modem

3 HART

®

signal

4 Other HART

®

devices or this signal converter (see also 7)

5 Signal converter terminals A (C)
6 Signal converter terminals A- (C-)
7 Signal converter with address > 0 and passive current output, connection of max. 15 devices (slaves) with 4…20 mA
8 Secondary Master
9 Power supply
10 Load ≥ 250 Ω (Ohm)