Operation – KROHNE OPTISONIC 7300 EN User Manual

Page 73

background image

OPERATION

6

73

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

frequency output X

C2.…

frequency output X

X stands for one of the connection terminals A, B or D.

…

stands for A, B or D.

C2.….1

pulse shape

Specify the pulse shape.

Select:

Symmetric (about 50% on and 50% off).

Automatic (constant pulse width, at 100% pulse rate about 50% on and

50% off)

Fixed (fixed pulse rate, setting see Fct. C2.….3 100% pulse rate)

C2.….2

pulse width

Only available if Fct. C2.….1 is set to "Fixed".

Range: 0.05…2000 ms

Note: max. setting value T

p

[ms] ≤ 500 / max. pulse rate [1/s], gives the

pulse width = time where the output is activated.

C2.….3

100 % pulse rate

Pulse rate for 100% of the measuring range.

Range: 0.0…10000 Hz

Limitation 100% pulse rate ≤ 100/s: I

max

≤ 100 mA

Limitation 100% pulse rate > 100/s: I

max

≤ 20 mA

C2.….4

measurement

Measurements for activating the output.

Select: volume flow / corr. volume flow / mass flow / molar mass / flow

speed / velocity of sound / gain / diagnostics 1, 2, 3

C2.….5

range

0…100% of the measurement set in Fct. C2.….4

0…xx.xx _ _ _ (format and unit depend on the measurement, see

above)

C2.….6

polarity

Set polarity of frequency output, please note flow direction in C1.3.2!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)

C2.….7

limitation

Set lower and upper limit for frequency output before applying the time

constant.

±

xxx … ±xxx%

Range: -150…+150%

C2.….8

low flow cutoff

Below set value, frequency output is set to zero.

x.xxx ± x.xxx%

Range: 0.0…20%

1

st

value = switching point / 2

nd

value = hysteresis

Condition: 2

nd

value ≤ 1

st

value

C2.….9

time constant

Averages measurement

Increasing value improves stability but delays the reaction.

Range: 000.1…100 s

C2.….10 invert signal

Define frequency output activation.

Off (switch closed)

On (switch open)