Operation – KROHNE OPTISONIC 7300 EN User Manual
Page 70
6
OPERATION
70
OPTISONIC 7300
www.krohne.com
09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en
C1.13
temperature
Only available if ‘fixed’ is selected in C1.12
Process temperature
Select value
Range: -40.00…+800.0 °C
C1.14
pressure
Only available if ‘fixed’ is selected in C1.12
Process pressure
Select value
Range: 1.00...250.00 bara
C1.15
density
Before entering first save the setting in C1.11 and leave the menu.
Select value for density at reference conditions as selected in C1.11
C1.16
diagnosis value
C1.16.1
diagnostics 1
Select: none / SNR 1 / gain 1 / velocity of sound 1
C1.16.2
diagnostics 2
Select: none / SNR 2 / gain 2 / velocity of sound 2
C1.16.3
diagnostics 3
Select: none / SNR 3 / gain 3 / velocity of sound 3
C2 inputs/outputs (I/Os)
C2.1
hardware
Assignment of connection terminals dependent on signal converter
version: active / passive / NAMUR
C2.1.1
terminal A
Select: off (switched off) / current output / frequency output / pulse
output / status output / limit switch / control input / current input
(for pressure)
C2.1.2
terminal B
Select: off (switched off) / current output / frequency output / pulse
output / status output / limit switch / control input / current input
(for temperature)
C2.1.3
terminal C
Select: off (switched off) / current output / status output / limit switch
C2.1.4
terminal D
Select: off (switched off) / frequency output / pulse output /
status output / limit switch
current input X
C2.
current input X
Only available if terminal A and B are current input.
X stands for one of the connection terminals A or B.
stands for A or B.
C2. .1
range 0%…100%
Current range for the selected measured variable, e.g. 4...20 mA,
corresponds to 0…100%.
xx.x … xx.x mA
Range: 04.0…20.0 mA
Condition: 4 mA ≤ 1
st
value ≤ 2
nd
value ≤ 20 mA
C2. .2
extended range
Exceeding the min. and max. limits.
xx.x … xx.x mA
Range: 00.5…23 mA
Condition: 0.5 mA ≤ 1
st
value ≤ 2
nd
value ≤ 23 mA
C2. .3
measurement
Terminal A: pressure
Terminal B: temperature