TECHNICAL DATA

8

93

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

Pressure

Pressure

Pressure

Pressure

All sensor designs at full rating acc. to below flange standards for standard

materials.

Max. pressure limited by

transducer

Titanium S7.01: 150 bara

Titanium S7.04: 101 bara

EN 1092-1

DN200…600: PN 10

DN100…150: PN 16

DN50...80: PN 40

ASME B16.5

2…24”: 150 lb RF

2…24”: 300 lb RF

2…24”: 600 lb RF

2...14": 900 lb RF

Higher pressure ratings on request

Properties of medium

Properties of medium

Properties of medium

Properties of medium (Other properties on request)

Physical condition

Dry gas

Density

Standard

Standard

Standard

Standard

10…45 g/mol / 1...150 kg/m

3

/ 0.062...9.36 lb/ft

3

Extended

Extended

Extended

Extended (may impose limitations on other spec's)

2…80 g/mol / 0.2...250 kg/m

3

/ 0.012...15.6 lb/ft

3

Installation conditions

Installation

For detailed information refer to

Installation

on page 16.

Inlet run

≤DN80: ≥ 20 DN
≥DN100: ≥ 10 DN

Outlet run

≥ 3 DN

Dimensions and weights For detailed information refer to

Dimensions and weights

on page 102.

Materials

Sensor

Sensor

Sensor

Sensor

NACE conformity

For standard range, all wetted materials are conform NACE MR175/103.

Flanges

(wetted)

Standard: carbon steel ASTM A105 N

Option: stainless steel 316 L, Carbon steel A350 LF2

Other materials on request.

Tube

(wetted)

Standard: carbon steel ASTM A106 Gr. B or Equivalent

Option: stainless steel 316 L, Carbon steel A333 GR6

Other materials on request.

Nozzles transducer

holders (wetted)

Stainless steel 316 Ti (1.4571)

Transducer holders

(wetted)

Stainless steel 316 L (1.4404)

Transducers (wetted)

Titanium grade 29

Transducer o-rings

(wetted)

Standard: FKM / FPM

Option: FFKM

Coating

Polyurethane