Technical data – KROHNE OPTISONIC 7300 EN User Manual
Page 93
TECHNICAL DATA
8
93
OPTISONIC 7300
www.krohne.com
09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en
Pressure
Pressure
Pressure
Pressure
All sensor designs at full rating acc. to below flange standards for standard
materials.
Max. pressure limited by
transducer
Titanium S7.01: 150 bara
Titanium S7.04: 101 bara
EN 1092-1
DN200…600: PN 10
DN100…150: PN 16
DN50...80: PN 40
ASME B16.5
2…24”: 150 lb RF
2…24”: 300 lb RF
2…24”: 600 lb RF
2...14": 900 lb RF
Higher pressure ratings on request
Properties of medium
Properties of medium
Properties of medium
Properties of medium (Other properties on request)
Physical condition
Dry gas
Density
Standard
Standard
Standard
Standard
10…45 g/mol / 1...150 kg/m
3
/ 0.062...9.36 lb/ft
3
Extended
Extended
Extended
Extended (may impose limitations on other spec's)
2…80 g/mol / 0.2...250 kg/m
3
/ 0.012...15.6 lb/ft
3
Installation conditions
Installation
For detailed information refer to
Installation
on page 16.
Inlet run
≤DN80: ≥ 20 DN
≥DN100: ≥ 10 DN
Outlet run
≥ 3 DN
Dimensions and weights For detailed information refer to
Dimensions and weights
on page 102.
Materials
Sensor
Sensor
Sensor
Sensor
NACE conformity
For standard range, all wetted materials are conform NACE MR175/103.
Flanges
(wetted)
Standard: carbon steel ASTM A105 N
Option: stainless steel 316 L, Carbon steel A350 LF2
Other materials on request.
Tube
(wetted)
Standard: carbon steel ASTM A106 Gr. B or Equivalent
Option: stainless steel 316 L, Carbon steel A333 GR6
Other materials on request.
Nozzles transducer
holders (wetted)
Stainless steel 316 Ti (1.4571)
Transducer holders
(wetted)
Stainless steel 316 L (1.4404)
Transducers (wetted)
Titanium grade 29
Transducer o-rings
(wetted)
Standard: FKM / FPM
Option: FFKM
Coating
Polyurethane