ELECTRICAL CONNECTIONS

4

49

OPTISONIC 7300

www.krohne.com

09/2012 - 4001102303 - MA OPTISONIC 7300 R03 en

4.7.5 Ex i inputs/outputs

Current output active (only current output terminals C/C- have HART

®

capability), Ex i

I/Os

• Observe connection polarity.
• U

int, nom

= 20 VDC

• I ≤ 22 mA
• R

L

≤ 450 Ω

• X designates the connection terminals A or C, depending on the version of the signal

converter.

Current output passive (only current output terminals C/C- have HART

®

capability),

Ex i I/Os

• Any connection polarity.
• U

ext

≤ 32 VDC

• I ≤ 22 mA
• U

0

≥ 4 V

• R

L, max

= (U

ext

- U

0

/ I

max

• X designates the connection terminals A or C, depending on the version of the signal

converter.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

INFORMATION!
For further information on electrical connection refer to Description of the inputs and outputs on
page 31

.

Figure 4-24: Current output active I

a

Exi

Figure 4-25: Current output passive I

p

Exi