Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 73
IM 11M13A01-04E
7-9
7. Startup
(2) Entering numeric values such as oxygen concentration values and factors
T7.4.5.2E.EPS
The set value is displayed after the parameter code selection.
An example of entering "9.8" is given below. (The currently set value is 0.0)
Touch the [>] key to move the position of the digit that is flashing to the
digit to be changed.
Continuously touch the [>] key, and the position of the digit that is flashing
will move continuously to the right.
Touch the [ ] key to set the numeric value 9.
Continuously touch the [ ] key, and the numeric value will change in sequence
from 0 to 1 to 2 to 3 to 8 to 9 and back to 0.
Touch the [>] key to move the position of the digit that is flashing to the right.
Touch the [ ] key to set the numeric value 8.
When the correct numeric value is displayed, touch the [ENT] key.
If you touch the [
ENT
] key again, the flashing stops and the current set value
will be in effect.
Touch the [
ENT
] key once again to return to the parameter code selection display.
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Switch operation
Display
Description
0
0.0
0
0
.0
0
9
.0
09.
0
09.
8
09.8
09.8
C
11
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
Ͼ
Ͼ
Ͼ
(3) If invalid numeric values are entered.
ENT
98.0
If an invalid numeric value (beyond the input range specified) is entered,
"ERR" will appear for twoseconds after touching the [
ENT
] key.
ENT
Err
ENT
0
0.0
"
ERR" appears for two seconds, and the display returns to the first set value.
Re-enter the numeric value.
T7.4.5.3E.EPS
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ
Ͼ