2 operational data initialization, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 126

IM 11M13A01-04E
10-8
10.2 Operational Data Initialization
Individual set data initialization enables you to return to the default values set at the time
of delivery. There are two types of initializations: an all set-data initialization and a
parameter-code-based initialization. Table 10.4 lists the initialization items by a param-
eter code, and default values.
Table 10.4 Parameter Codes for Initialization
Parameter code
Data to be initialized
F30
All data
F31
Data in Group A
F32
Data in Group B
F33
Data in Group C
F34
Data in Group D
F35
Data in Group E
F36
Data in Group F
T10.4EPS
CAUTION
When Data in Group Fare initialized by the parameter code of F36, F01 and F02 and
F08 and F10 cannot be initialized.