8. Detailed Data Setting
8.7.2
Setting Periods over which Average Values are Calculated and Periods over which
Maximum and Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maxi-
mum and minimum values under measurement (see Section 10.1.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration
average values are calculated and maximum and minimum values are monitored.
8.7.2.1 Procedure
Use the parameter-code table below to set the average, maximum and minimum oxygen
concentration values. Periods over which average is calculated and periods over which
maximum and minimum values are monitored can be set, ranging from 1 to 255 hours.
If the set ranges are beyond the limits specified, an “ERR” will be displayed.
Table 8.15 Parameter Codes for Average, Maximum and Minimum Values
Set item
Parameter code Set range Units
Periods over which average
F11
1 to 255
Hours
values are calculated
Periods over which maximum
F12
1 to 255
Hours
and minimum values are
monitored
T8.15E.EPS
8.7.2.2 Default Value
When the analyzer is delivered, or if data are initialized, periods over which average
values are calculated are set to one hour, and periods over which maximum and mini-
mum values are monitored are set to 24 hours.