14 output-related items in group c – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 132

IM 11M13A01-04E
10-14
Output-related Items in Group C
Code
Item
Tuning
Engineering Default setting
unit
C01
Analog output
0: Oxygen concentration
Oxygen concentration
1: Amount of moisture content
2: Mixed ratio
C03
Output mode
0: Linear
Linear
1: Logarithm
C04
Output during warm-up 0: Held at 4 mA
up
1: Held at 20 mA
2: Set value remains held.
Held at 4 mA.
C05
Output during
0: Not held
maintenance
1: Held output just before
Held output just before
maintenance service.
maintenance service.
2: Set value remains held.
C06
Output during
0: Not held
Held output just before
calibration 1: Held output just before
calibration.
calibration.
2: Set value remains held.
C07
Output in abnormal
0: Not held
Held output at a preset
state
1: Held output just before
value.
abnormal state occurs.
2: Set value remains held.
C11
Min. oxygen
See Section 8.2.
% O
2
0% O
2
concentration
C12
Max. oxygen
See Section 8.2.
% O
2
25% O
2
concentration
C13
C14
C15
C16
C30
Output damping
0 to 255
Seconds
0 second
factor
C31
Set value during
2.4 to 21.6
mA
4 mA
warm-up
C32
Set value during
2.4 to 21.6
mA
4 mA
maintenance
C33
Set value during
2.4 to 21.6
mA
4 mA
calibration
C34
Set value in abnormal 2.4 to 21.6
mA
3.4 mA
state
T10.6C.EPS