Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 135

IM 11M13A01-04E
10-17
10. Other Functions
Equipment Setup and Others in Group F
T10.6F.EPS
Code Item
Tuning
Engineering Default setting
unit
F01 Equipment setup
0: Oxygen analyzer
Oxygen analyzer
1: Humidity analyzer
F02 Selection of measurement gas 0: Wet
Wet
1: Dry
F04 Selection of temperature units
0: degree C
1: degree F
degree C
F05 Selection of pressure units
0: kPa
1: psi
kPa
F08 Selection of display items
0: Oxygen concentration
Oxygen concentration
1: Amount of moisture
quantity
2: Mixed ratio
3: Item selected with
analog output
F10 Date
YY.MM.DD/H
H.MM
F11 Period over which average
1 to 255 hours
Hours
One hour
values are calculated
F12 Period over which max. and
1 to 255 hours
Hours
24 hours
min. values are monitored
F13
F14
F20 Content of moisture in exhaust
0 to 5
m
3
/kg (m
3
) 1.0 m
3
/kg (m
3
)
gas
F21 Theoretical air quantity
0 to 20
m
3
/kg (m
3
) 1.0 m
3
/kg (m
3
)
F22 X value
0 to 19.99
1.0
F23 Ambient absolute
0 to 1
kg/kg
0.1 kg/kg
humidity
F30 Initializing all data
F31 Initializing data in group A
F32 Initializing data in group B
F33 Initializing data in group C
F34 Initializing data in group D
F35 Initializing data in group E
F36 Initializing data in group F