6 input contact settings, Caution, 1 setting input contact – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 101: 2 default values

IM 11M13A01-04E
8-13
8. Detailed Data Setting
8.6
Input Contact Settings
The converter input contacts execute set functions by accepting a remote (contact)
signal. Table 8.12 shows the functions executed by a remote contact signal.
Table 8.12 Input Contact Functions
Set item
Function
Calibration-gas drop
While a contact signal is on, neither semi-automatic nor
automatic calibrations can be made.
Measuring range change
While contact input is On, the analog output range is switched to
0-25% O
2
.
Calibration start
If a contact signal is applied, semi-automatic calibration starts
(only if the semi-automatic or automatic mode has been setup).
Calibration starts with an applied one-second time interval
single-output contact signal. Even though a continuous contact
signal is applied, a second calibration cannot be made. If you
want to make a second calibration, turn the contact signal off
and then back on.
Unburnt gas detection
If a contact signal is on, the heater power will be switched off.
(A one-second time interval single-output signal is available as
a contact signal.) If this operation starts, the sensor
temperature decreases and an error occurs. To restore it to
normal, turn the power off and then back on, or reset the
analyzer.
T8.12E.EPS
CAUTION
To conduct a semi-automatic calibration, be sure to set the Calibration setup mode to
“Semi-Auto” or “Auto”.
8.6.1
Setting Input Contact
To set the input contacts, follow the parameter codes given in Table 8.13.
Table 8.13 Parameter Codes for Input Contact Settings
Set item
Parameter code
Set value
Input contact 1 (function)
E01
0: Invalid
1: Calibration gas-pressure decrease
2: Measuring range change
3: Calibration
4: Unburnt gas detection
Input contact 2 (function)
E02
0: Invalid
1: Calibration gas-pressure decrease
2: Measuring range change
3: Calibration
4: Unburnt gas detection
Input contact 1 (action)
E03
0: Operated when closed
1: Operated when open
Input contact 2 (action)
E04
0: Operated when closed
1: Operated when open
T8.13E.EPS
8.6.2
Default Values
When the analyzer is delivered, or if data are initialized, the input contacts are all open.