Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 22

IM 11M13A01-04E
2-4
FM Explosionproof: ZR202S-B
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure Rating: NEMA 4X
Temperature Class: T2
CSA Explosionproof: ZR202S-C
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
Enclosure: Type 4X
Temperature Class: T2
IECEx Flameproof: ZR202S-D
IECEx flameproof type
Ex d IIB + H
2
T2
Enclosure: IP66
IECEx type of protection “Dust”
Ex tD A21 IP66 T300
ЊC
Enclosure: IP66
NAME PLATE
INTEGRAL TYPE
ZIRCONIA ANALYZER
MODEL
SUFFIX
ZR202S
NO.
AMB.TEMP. -20 TO 55
Њ
C
STYLE
SUPPLY
100-240VAC
50/60Hz MAX.300W
Tokyo 180-8750 JAPAN
MODEL : Specified model code
SUFFIX : Specified suffix code
STYLE : Style code
AMB. TEMP : Ambient temperature
NO. : Serial No. and year of production*1
YOKOGAWA
᭜
: The manufacturer name
Tokyo 180-8750 JAPAN : The manufacturer address*2
*1: The third to seventh figure from the last shows the year of production.
e.g. 27D327560 2005.02
*2: "180-8750" is a zip code which represents the following address.
2-9-32 Nakacho, Musashino-shi, Tokyo Japan
No. IECEx KEM 06.0006
Ex d IIB
ϩ
H
2
T2
Ex tD A21 IP66 T300
Њ
C
ENCLOSURE: IP66
The country of origin.
Can be operated in the field without opening the cover using optical switches.
Display: 6-digit LCD
Switch: Three optical switches
Output Signal: 4 to 20 mA DC, one point (maximum load resistance 550
⍀)
Digital Communication (HART): 250 to 550
⍀, depending on quantity of field
devices connected to the loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Contact Output Signal: Two points (one is fail-safe, normally open)
Contact Input Signal: Two points
Sample Gas Temperature: 0 to 700
ЊC
It is necessary to mount the cell using Inconel cell-bolts when the tem-
perature measures more than 600
ЊC.
High-temperature service—greater than 700
ЊC—is not available.