A1-4 – Yokogawa EJX930A User Manual
Page 81
A1-4
IM 01C25T02-01E
Example of the case where SWAP_2 is on (monotone increase):
The input range of IN_1 is always in CURVE_X. The following shows the input/output graph of the IN_1 values.
Y
Output
Y6
(High limit)
Y1
(Low limit)
X1
X2
X3
X4
X5
X6
X7 =INFINITY
X
Input
FA0104.ai
Figure A1.4 Example of Curve for IN_1 (SWAP_2 = on)
The input range of IN_2 is always in CURVE_Y. The following shows the input/output graph of the IN_2 values.
Y1
(Low limit)
X6
X5
X4
X3
X2
X1
Y2 Y3
Y4
Y5 Y6 Y7 =INFINITY
Y Input
FA0105.ai
(High limit)
Output
X
Figure A1.5 Example of Curve for IN_2 (SWAP_2 = on)
When SWAP_2 is on, the array elements of CURVE_Y must be configured for a monotone increase or
decrease. (Y1 < Y2 < Y3 < Y4 < Y5 < Y6 or Y6 < Y5 < Y4 < Y3 < Y2 < Y1)