2 sensor transducer block, Sensor transducer block -5 – Yokogawa EJX930A User Manual
Page 69
<9. Parameter Lists>
9-5
IM 01C25T02-01E
Relative
Index Index
Parameter Name
Factory Default
Write
Mode
Explanation
119
1119
DEVICE_
CONDITION_
ACTIVE_4
0x00000000
—
FD_EXTENDED_ACTIVE_4 & FD_EXTENDED_MAP_4
120
1120
DEVICE_
CONDITION_
ACTIVE_5
0x00000000
—
FD_EXTENDED_ACTIVE_5 & FD_EXTENDED_MAP_5
121
1121
DEVICE_
CONDITION_
ACTIVE_6
0x00000000
—
FD_EXTENDED_ACTIVE_6 & FD_EXTENDED_MAP_6
122
1122
DEVICE_
CONDITION_
ACTIVE_7
0x00000000
—
FD_EXTENDED_ACTIVE_7 & FD_EXTENDED_MAP_7
123
1123
DEVICE_
CONDITION_
ACTIVE_8
0x00000000
—
FD_EXTENDED_ACTIVE_8 & FD_EXTENDED_MAP_8
9.2 SENSOR Transducer Block
Relative
Index Index
Parameter Name
Factory Default Write
Mode
Explanation
0
2000
Block Header
TAG: “STB”
Block
Tag =
O/S
Information on this block such as Block Tag, DD Revision,
Execution Time etc.
1
2001
ST_REV
—
—
The revision level of the static data associated with the function
block. The revision value will be incremented each time a static
parameter value in the block is changed.
2
2002
TAG_DESC
Null
AUTO The user description of the intended application of the block
3
2003
STRATEGY
1
AUTO The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block.
4
2004
ALERT_KEY
1
AUTO The identification number of the plant unit. This information
may be used in the host for sorting alarms, etc.
5
2005
MODE_BLK
AUTO
AUTO The actual, target, permitted, and normal modes of the block.
6
2006
BLOCK_ERR
—
—
This parameter reflects the error status associated with
hardware or software components associated with a block. It is
a bit string, so that multiple errors may be shown.
7
2007
UPDATE_EVT
—
—
This alert is generated by any change to the static data.
8
2008
BLOCK_ALM
—
—
The block alarm is used for all configuration, hardware,
connection failure or system problems in the block. The cause
of the alert is entered in the subcode field. The first alert to
become active will set the Active status in the Status attribute.
9
2009
TRANSDUCER_
DIRECTORY
—
—
A directory that specifies the number and starting indices of the
transducers.
10
2010
TRANSDUCER_
TYPE
100 (Standard
Pre-ssure with
Calibration)
—
Identifies transducer.
11
2011
XD_ERROR
—
—
The error code in transducer.
0=No failure
18=Calibration error
20=Electronics failure
22=I/O failure
12
2012
COLLECTION_
DIRECTORY
—
—
A directory that specifies the number, starting indices, and DD
Item Ids of the data collections in each transducer within a
transducer block.
13
2013
PRIMARY_ VALUE_
TYPE
107: differential
pressure
108: gauge
pressure
109: absolute
pressure
O/S
The type of measurement represented by primary value.
Followings are available for the transmitter:
107=differential pressure
108=gauge pressure
109=absolute pressure
14
2014
PRIMARY_ VALUE
—
—
The measured value and status available to the function block.
15
2015
PRIMARY_ VALUE_
RANGE
Range of capsule —
The High and Low range limit values, engineering units code
and the number of digits to the right of the decimal point to be
used to display the primary value.
16
2016
CAL_POINT_HI
Max range
O/S
The highest calibrated value.
17
2017
CAL_POINT_LO
0
O/S
The lowest calibrated value.