4 iecex certification, Iecex certification -12 – Yokogawa EJX930A User Manual
Page 23
<2. Handling Cautions>
2-12
IM 01C25T02-01E
2.1.4 IECEx Certification
a. IECEx Flameproof Type
Caution for IECEx flameproof type.
Note 1. The transmitters with optional code /SF2
are applicable for use in hazardous
locations:
• No. IECEx CSA 07.0008
• Applicable Standard:
IEC60079-0:2004, IEC60079-1:2003
• Flameproof for Zone 1, Ex d IIC T6...T4
• Enclosure: IP66 and IP67
• Maximum Process Temperature:
120°C (T4), 100°C (T5), 85°C (T6)
• Ambient Temperature:
–50* to 75°C (T4), –50* to 80°C (T5),
–50* to 75°C (T6)
* –15°C when /HE is specified.
• Supply Voltage: 32 V dc max.
• Output Signal: 15 mA dc
Note 2. Wiring
• In hazardous locations, the cable entry
devices shall be of a certified flameproof
type, suitable for the conditions of use and
correctly installed.
• Unused apertures shall be closed with
suitable flameproof certified blanking
elements.
Note 3. Operation
• WARNING:
AFTER DE-ENERGIZING, DELAY 5
MINUTES BEFORE OPENING.
• WARNING:
WHEN AMBIENT TEMPERATURE ≥ 65°C,
USE THE HEAT-RESISTING CABLES ≥
90°C.
• Take care not to generate mechanical
sparking when accessing to the instrument
and peripheral devices in a hazardous
location.
Note 4. Maintenance and Repair
• The instrument modification or parts
replacement by other than authorized
representative of Yokogawa Electric
Corporation is prohibited and will void IECEx
Certification.
b. IECEx Intrinsically Safe Type / type n
Caution for IECEx Intrinsically safe and type n.
Note 1. The transmitters with optional code /SS25
are applicable for use in hazardous
locations
• No. IECEx CSA 05.0012
• Applicable Standard:
IEC 60079-0:2000, IEC 60079-11:1999,
IEC 60079-15:2001
• Ex ia IIC/IIB T4, Ex nL IIC T4
• Ambient Temperature: –40* to 60°C
* –15°C when /HE is specified.
• Max. Process Temp.: 120°C
• Enclosure: IP66 and IP67
Note 2. Entity Parameters
• Intrinsically safe ratings are as follows:
[Entity]
Maximum Input Voltage (Vmax/Ui) = 24 V
Maximum Input Current (Imax/Ii) = 250 mA
Maximum Input Power (Pmax/Pi) = 1.2 W
Maximum Internal Capacitance (Ci)
= 1.76 nF
Maximum Internal Inductance (Li) = 0 µH
[FISCO IIC]
Maximum Input Voltage (Vmax/Ui) = 17.5 V
Maximum Input Current (Imax/Ii) = 380 mA
Maximum Input Power (Pmax/Pi) = 5.32 W
Maximum Internal Capacitance (Ci)
= 1.76 nF
Maximum Internal Inductance (Li) = 0 µH
[FISCO IIB]
Maximum Input Voltage (Vmax/Ui) = 17.5 V
Maximum Input Current (Imax/Ii) = 460 mA
Maximum Input Power (Pmax/Pi) = 5.32 W
Maximum Internal Capacitance (Ci)
= 1.76 nF
Maximum Internal Inductance (Li) = 0 µH
• Type "n" ratings are as follows:
Maximum Input Voltage (Vmax/Ui) = 32 V
Maximum Internal Capacitance (Ci)
= 1.76 nF
Maximum Internal Inductance (Li) = 0 µH
• Installation Requirements
Uo ≤ Ui, Io ≤ Ii, Po ≤ Pi,
Co ≥ Ci + Ccable, Lo ≥ Li + Lcable
Voc ≤ Vmax, Isc ≤ Imax,
Ca ≥ Ci + Ccable, La ≥ Li + Lcable
Uo, Io, Po, Co, Lo, Voc, Isc, Ca and La are
parameters of barrier.