A8-20 – Yokogawa EJX930A User Manual
Page 164
A8-20
IM 01C25T02-01E
Checklist (4/5)
No.
Items
Parameters
Result
Example
10-a-2 Low Side Blocking
• Close the low-pressure side valve
completely.
• Record the values of VALUE_# after
the certain time, (DIAG_PERIOD ×
DIAG_COUNT), passed.
*: Record the value after checked that
the status is “GOOD”.
VALUE_FDP (Index: 2086) *
7.48562e-9
VALUE_FSPL (Index: 2087) *
7.23277e-9
VALUE_FSPH (Index: 2088)
7.14085e-9
VALUE_BLKF (Index: 2089) *
-0.287259
• Record the status of Checkbox in
DIAG_OPTION.
• Check that the alarms status of “A
Blocking” and “Low Side Blocking” are
set.
Note: If the alarm of “Outside Diagnosis
Range” is generated, the valve may
be closed too much tightly. Open
valve a little and record the updated
status of the parameters.
DIAG_OPTION (Index: 2072)
A Blocking
Large Fluctuation of Low Side
Large Fluctuation of High Side
Low Side Blocking
High Side Blocking
B Blocking
Invalid Ref BlkF
Invalid Ref fSPh
Invalid Ref fSPl
Invalid Ref fDP
Outside Diagnosis Range
Reflect Blockage to PV/SV/TV
Status
• Check that the alarm of “Low Side
Blocking” is generated.
• Check that the alarm of “High Side
Blocking” is not generated.
DIAG_ERR (Index: 2069)
Low Side Blocking
High Side Blocking
10-a-3 Both Side Blocking
• Close the both-pressure side valves
completely.
• Record the values of VALUE_# after
the certain time, (DIAG_PERIOD ×
DIAG_COUNT), passed.
*: Record the value after checked that
the status is “GOOD”.
VALUE_FDP (Index: 2086)
7.48562e-9
VALUE_FSPL (Index: 2087)
7.23277e-9
VALUE_FSPH (Index: 2088)
7.14085e-9
VALUE_BLKF (Index: 2089)
-0.287259
• Record the status of Checkbox in
DIAG_OPTION.
• Check that the alarms status of “High
Side Blocking” , “Low Side Blocking”
and “B Blocking” are set.
Note: If the alarm of “Outside Diagnosis
Range” is generated, the valve may
be closed too much tightly.
Open valve a little and record the
updated status of the parameters.
DIAG_OPTION (Index: 2072)
A Blocking
Large Fluctuation of Low Side
Large Fluctuation of High Side
Low Side Blocking
High Side Blocking
B Blocking
Invalid Ref BlkF
Invalid Ref fSPh
Invalid Ref fSPl
Invalid Ref fDP
Outside Diagnosis Range
Reflect Blockage to PV/SV/TV
Status
• Check that the alarm of “B Blocking” is
generated.
DIAG_ERR (Index: 2069)
B Blocking