Deskew, Enhancedresolution – Teledyne LeCroy WaveRunner Automation Command User Manual
Page 336
Automation Command and Query Reference Manual - Processor Reference
Sets/Queries whether the autoscale function is enabled for the derivative function trace Fx. If enabled,
an auto-scale operation is performed whenever the setup changes.
Description
FindScale
Action
Initiates a Find Scale action, to set a suitable vertical scale for the derivative function trace Fx.
Description
VerOffset
Double
From -1e+006 to 1e+006 step 1e-009
Range
Sets/Queries the vertical offset of the derivative function trace Fx.
Description
VerScale
DoubleLockstep
From 1e-012 to 1e+013 step 10000, locked to 1 2 5, fine grain allowed=false, on=false
Range
Sets/Queries the vertical scale of the derivative function Fx.
Description
DESKEW
app.Math.Fx.OperatorYSetup (Operator = "Deskew")
Deskew waveform by shifting it in time.
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Set the displacement of the trace F3 to 3.7e-9
app.Math.F3.View = True
app.Math.F3.MathMode = "OneOperator"
app.Math.F3.Operator1 = "Deskew"
app.Math.F3.Operator1Setup.WaveDeskew
Example
WaveDeskew
Double
From -2.5e-008 to 2.5e-008 step 1e-012
Range
Sets/Queries the displacement in time of the trace Fx. A positive value delays the signal: a negative
one makes it appear earlier.
Description
WaveDeskew
Double
ENHANCEDRESOLUTION
app.Math.Fx.OperatorYSetup (Operator = "EnhancedResolution")
Bits
Enum
Bits
Enum
2-5
918500 RevA