Scanoverlay – Teledyne LeCroy WaveRunner Automation Command User Manual
Page 315

Automation Command and Query Reference Manual - Control Reference
VariableVerZoom
Bool
Enable/Disable the variable Vertical Zoom control. If enabled, the VerZoom control may be set to a
value other than the standard 1, 2, 5 sequence.
Description
VerPos
Double
From -1.5 to 1.5 step (8 digits)
Range
Vertical Position of the trace, normalized to a value between -1.5 and 1.5. A value of zero is the
default, and indicates no position change relative to the source trace.
Description
VerZoom
Double
From 0.1 to 100 step (8 digits)
Range
Vertical Zoom setting. Locked to a 1, 2, 5 sequence unless VariableVerZoom is set to True .
Description
SCANOVERLAY
app.WaveScan.ScanOverlay
AxisXRotation
Integer
From -90 to 90 step 1
Range
AxisYRotation
Integer
From -90 to 90 step 1
Range
This control is used only when Persisted is true and Persistence3d is true. It controls rotation about
the Y axis of the view being persisted.
Description
ClearSweeps
Action
Clear any accumulated result data. Useful for example to restart an average, or parameter statistics.
Description
AxisXRotation
Integer
AxisYRotation
Integer
ClearSweeps
Action
EnablePersistence
Bool
Persist3DQuality
Enum
Persisted
Bool
Persistence3d
Bool
PersistenceMonoChrome
Bool
PersistenceSaturation
Integer
PersistenceTime
Enum
ShowLastTrace
Bool
View
Bool
1-307
918500 RevA