Result, Tief – Teledyne LeCroy WaveRunner Automation Command User Manual
Page 231

Automation Command and Query Reference Manual - Control Reference
TIESlope
Enum
Values
Both
Neg
Pos
UseAllEdges
Bool
View
Bool
Sets/Queries the visibility of the function.
Description
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Show the TIE function.
app.SDA.TIE.View = True
Example
RESULT
app.SDA.TIE.Out.Result
Properties of the type xxxx.Out.Result.yyyy are those of the last completed acquisition. They are not affected if other
cvars are changed after that acquisition was completed. This distinction between "Out.Result" properties and other
cvars is most important when the trigger mode is Single or Stopped. You should treat "Out.Result" properties as read-
only.
TIEF
app.SDA.TIEF
BitRate
Double
From 90000 to 4e+010 step 1
Range
ClearSweeps
Action
Clear any accumulated result data. Useful for example to restart an average, or parameter statistics.
Description
View
Bool
Sets/Queries the trace's 'Viewed' state. When true, the trace is displayed on one of the display
graticules. Note that even when a trace is not visible, it may be used as a source for Math, Measure,
etc.
Description
BitRate
Double
ClearSweeps
Action
View
Bool
1-223