Teledyne LeCroy WaveRunner Automation Command User Manual
Page 297

Automation Command and Query Reference Manual - Control Reference
MarkerToCenterFreq
Action
Center the spectrum on the current market frequency.
Description
MaxFrequency
Double
From 100000 to 1e+011 step 1
Range
Read-only control, reflects the maximum frequency of the spectrum. Calculated as 1/2 the scope
sample rate.
Description
MaxPeaks
Integer
From 1 to 100 step 1
Range
Define the maximum number of peaks which will be measured, and presented in the table.
Description
Mode
Enum
Define the spectrum analyzer mode.
Description
Values
Average
Average a number of traces, specified in the NumAverages control.
MaxHold
Record the maximum value in each frequency bin.
Normal
Normal Mode.
ReferenceFreq
Double
From 0 to 1e+010 step 1
Range
Defines the reference frequency, at which point a vertical marker (cursor) will be displayed.
Description
ReferenceLevel
Double
From -200 to 200 step 0.0001
Range
Define the vertical reference level, that is the level in dBm, that is shown at the top of the graticule.
Description
ResolutionBandwidth
DoubleLockstep
From 0.1 to 1e+009 step 1000, locked to 1 3 5, fine grain allowed=false, on=false
Range
When not in Auto Resolution Bandwidth mode, this control is used to request a specific resolution
bandwidth.
Note that the requested resolution bandwidth is not always achieveable, so a second control,
ActualResolutionBandwidth is available to verify that the actual resolution bandwidth is acceptable.
Description
1-289
918500 RevA