beautypg.com

Teledyne LeCroy WaveRunner Automation Command User Manual

Page 16

background image

Automation Command and Query Reference Manual - Control Reference

Calibrate

Action

Initiates a full calibration of the acquisition system of the instrument.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Start a calibration.
app.Acquisition.Calibrate

Example

CalibrateAll

Action

CalNeeded

Integer

From -2147483648 to 2147483647 step 1

Range

Query: Indicates whether calibration is required or not.

Based on hexadecimal bit value, it provides following information:
0x00000001: Front end calibration is required
0x00000002: Digitizers delay matching is required
0x00000004: Digitizers gain matching is required
0x00000008: Trigger level calibration is required
0xFFFFFFFF(-1): All of above calibrations are required

Description

ClearSweeps

Action

Resets any accumulated average data or persistence data for channel waveforms (C1..C4). Valid only
when one or more channels have waveform averaging or persistence enabled in their pre-processing
settings. Note that an average may be reset on an individual basis using
app.Acquisition.Cx.ClearSweeps control.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Clear accumulated sweeps for channels C1...C4
app.Acquisition.ClearSweeps

' Clear accumulated sweeps for only C1
app.Acquisition.C1.ClearSweeps

Example

HorOffset

Double

From -2.5e-007 to 2.5e-007 step 1e-009

Range

same as "app.Acquisition.Horizontal.HorOffset.cvar"

Description

1-8

918500 RevA