Section 5: Serial I/O

04/14

54

51216 -- Rev. 6

5.4.8. NexWeigh Serial Commands and Definitions,

Continued

kk)

Set Date

Definition: Used to set the date in the RTC.

Command: SDA

Options:

mm/dd/yy

Sample Command: "SDA 05/28/99\r"

ll)

Set Security Level

Definition: Used to set the scale’s security level. Only valid levels between 0 and
MAXSECURITYLEVEL allowed. This can be a dangerous command. One could
cause themselves to be locked out from further commands.

Command: ‘SSL’ ‘level’

Options:

0-5

Sample Command: "SSL 1\r"

mm)

Set Capacity

Definition: Used to set the scale’s capacity in Primary units.

Command: ‘STC’ ‘value’

Options:

decimal weight value

Sample Command: "STC 100 \r"

nn)

Set Test Weight

Definition: Used to set the scale’s test weight value in Primary units.

Command: ‘STW’ ‘value’

Options:

decimal weight value

Sample Command: "STW 75.5 \r"

oo)

Set Accumulate Enable

Definition: Used to set the Accumulate enable parameter for the scale.

Command: ‘SAC’ ‘mode’

Options: E,D

Sample Command: "SAC E\r"

pp)

Set Output ID Enable

Definition: Used to set the Output Scale Id enable parameter for the scales serial
output stream.

Command: ‘SOI’ ‘mode’

Options:

E,D

Sample Command: "SOI E\r"