Fairbanks NexWeigh User Manual
Page 52

Section 5: Serial I/O
04/14
52
51216 -- Rev. 6
5.4.8. NexWeigh Serial Commands and Definitions,
Continued
z)
Set Flush Filter Band
Definition: Used to set scale’s flush filter band in divisions. Only valid settings allowed.
Command: ‘SFF’ ‘divisions’
Options:
1, 2, 5, 10, 20, 50, 100, 5000,
Sample Command: "SFF 5000 \r"
aa)
Set Display Intensity
Definition: Used to set scales display intensity parameter. Only valid settings allowed.
Command: ‘SDI’ ‘level’
Options:
1-7
Sample Command: "SDI 4\r"
bb)
Set Zero Limit
Definition: Used to set the scales zero limit parameter. 100% for USA or 2% for
Canada. Only valid settings allowed.
Command: ‘SZL’ ‘range’
Options:
100, 2
Sample Command: "SZL 100\r"
cc)
Set OverLoad Limit
Definition: Used to set the scales Over capacity limit parameter. Range is in %. Only
valid settings allowed.
Command: ‘SOL’ ‘range’
Options:
102.5, 105.0, 110.0, 150.0 etc.
Sample Command: "SOL 102.5\r"
dd)
Set UnderLoad Limit
Definition: Used to set the scales Under capacity limit parameter. Range is in %.
Only valid settings allowed.
Command: ‘SUL’ ‘range’
Options:
25, 50, 100, etc.
Sample Command: "SUL 25\r"
ee)
Set Custom Factor
Definition: Used to set the custom units conversion factor. Value is a number representing
conversion from primary unit to custom unit, i.e. 8.33 lbs per gallon of water.
Command: ‘SCF’ ‘value’
Options:
decimal value
Sample Command: "SCF 8.33 \r"