Configuration menu, continued tare – Fairbanks NexWeigh User Manual
Page 29
Section 4: Programming Configuration
04/14
29
51216 -- Rev. 6
4.5.2. Configuration Menu, Continued
tArE:
Programming the tare setting: (the means by which a container’s weight
may be removed, to set the instrument to display the net weight only) The display
will indicate
tArE
followed by the current tare setting. Press
OVER/UNDER/NEXT
,
UNITS
or
TARE
to change the selection. Press
PROGRAM/ENTER
to enter the selection. Available selections are:
OFF
Tare is disabled.
ON
Tare is active.
On-CLr
Tare automatically clears when Gross weight returns to Zero.
*
Underline identifies the program default
HL:
Overload limit: The actual value at which the display goes to
OL
(represented
as a percentage of capacity). The display will indicate
HL
followed by the current
setting. Press
OVER/UNDER/NEXT
or
UNITS
to scroll up, or
TARE
to scroll
down through the available choices. Press
PROGRAM/ENTER
to enter the
selection. Available selections are:
102.5
103
105
110
300
UL:
U
nderload
L
imit: The actual value at which the display goes to
UL
(represented as a percentage of capacity behind zero). The display will indicate
UL
followed by the current setting. Press
OVER/UNDER/NEXT
or
UNITS
to scroll
up, or
TARE
to scroll down through the available choices. Press
PROGRAM/ENTER
to enter the selection. Available selections are:
5.5
10
25
50
100