Setup menu, continued int, Configuration menu – Fairbanks NexWeigh User Manual
Page 25

Section 4: Programming Configuration
04/14
25
51216 -- Rev. 6
4.5.1. Setup Menu, Continued
Int:
Display brightness setting: This changes the intensity of the display brightness.
The display will indicate Int followed by the intensity setting from
1-7
, with
7
being
the brightest. Press
OVER/UNDER/NEXT
or
UNITS
to scroll up, or
TARE
to
scroll down through the available choices. Press
PROGRAM/ENTER
to enter the
selection. The default setting is
4
.
The NexWeigh instrument will return to the
SEtuP
prompt. Navigate to the desired
menu heading to continue programming, or select
Done
to finish.
4.5.2. Configuration Menu
4. Press and hold the
PROGRAM/ENTER
key or press
S1
on the main PCB to
display
SEtUP
.
5. Press
UNITS
or
OVER/UNDER/NEXT
to display ConFiG.
6. Press the
PROGRAM/ENTER
key to enter setup menu.
CAP:
Programming the scale capacity: The display will indicate
CAP
followed by
the current setting with the most significant digit blinking. Press
UNITS
to scroll up,
or the
1
to scroll down to the desired numeric value. The
OVER/UNDER/NEXT
key will move the flashing digit to the right, and the
TARE
key will move it left.
Press
PROGRAM/ENTER
when the display shows the correct capacity setting.
NOTE:
Capacity can be up to six (6) displayed digits.