Configuration menu, continued cust – Fairbanks NexWeigh User Manual
Page 27

Section 4: Programming Configuration
04/14
27
51216 -- Rev. 6
4.5.2. Configuration Menu, Continued
CuSt:
Custom units. The display will indicate CuSt and the current selection,
either
Y
for enable or
N
for disable. Press
OVER/UNDER/NEXT
,
UNITS
or
TARE
to change the selection. Press
PROGRAM/ENTER
to enter the selection.
NOTE:
When custom units are active, units LED instruments are off in the weigh
mode.
AZt:
Programming the Automatic Zero Tracking band: (this feature will maintain
zero when small amounts of material are placed on the scale, such as rain, snow,
debris, etc.) The display will indicate
AZt X
where
X
is the current setting. Press
OVER/UNDER/NEXT
,
UNITS
or
TARE
to change the selection. Press
PROGRAM/ENTER
to enter the selection. Available selections are:
OFF
Function is not active.
0.5
Half (½) of a division / increment / graduation.
1
One (1) division / increment / graduation.
3
Three (3) divisions / increments / graduations.
*
Underline identifies the program default.
bAL:
Programming the motion band (the range in divisions/increments/graduations
that weight must be stable before a print, zero, or tare function will be allowed). The
display will indicate
bAL X
where
X
is the current setting. Press
OVER/UNDER/NEXT
,
UNITS
or
TARE
to change the selection. Press
PROGRAM/ENTER
to enter the selection. Available selections are:
OFF
Function is not active.
0.5
Half (½) of a division / increment / graduation.
1
One (1) division / increment / graduation.
3
Three (3) divisions / increments / graduations.
*
Underline identifies the program default.