Luminex 100 IS Developer Workbench Guide Version 2.3 User Manual
Page 67

PN 89-00002-00-084 Rev. B
61
Compatibility with Luminex 100
Version 1.7
Certain settings in the Luminex 100 Version 1.7 software must have
particular values for your assay to be compatible with the Luminex
100 IS system. See Table 5.
You cannot have Luminex 100 Version 1.7 and Luminex 100 IS
loaded on the same system. Record your gate values on the Luminex
100 Version 1.7 system and then enter the information into templates
on the Luminex 100 IS system
To set the Luminex 100 Version 1.7 bead map to be compatible
with Luminex 100 IS Version 2.3:
1. Select the Luminex 100 Version 1.7 icon on your desktop. The
application loads and displays the opening screen.
2. Click Options on the toolbar. At the Options dialog box, User
tab, verify that the Default Bead Map is set to 100 regions. If it
is not, click Browse and select the 100 regions bead map. Click
OK. See Figure 40.
B
Table 5: Value Settings for the Luminex 100
Default bead map:
100 regions
Sample size:
between 20 and 200 µL
Events per bead:
recommend 100
Flow rate:
Fast