Enumerate the Coupled
Microspheres

1. Dilute the resuspended coupled microspheres 1:100 in deionized

water.

2. Mix thoroughly using a vortex.

3. Transfer 10 µL to a hemacytometer.

4. Count the microspheres within the 4 large corners of the

hemacytometer grid.

5. Microspheres/µL = (Sum of microspheres in 4 large corners) x

2.5 x 100 (dilution factor).

6. Store the preparation at 2

o

C - 8°C. Protect it from light.

Technical Notes

1. Synthesize the Oligonucleotides with a 5’ or 3’

amino group and

spacer modification. No Tris or Azide or other amine-containing
buffers should be present during the coupling procedure.
Preferably, resuspend the oligonucleotide in water following
synthesis. If oligonucleotides were previously solubilized in an
amine-containing buffer, then precipitation and resuspension into
water. Refer to http://luminexcorp.custhelp.com for more
information.

2. You can scale this procedure up or down. Refer to http://

luminexcorp.custhelp.com for more information.

3. The optimal coupling concentration for a given oligonucleotide

is determined by coupling at various concentrations within the
recommended range of 1 to 100 µM.

4. Minimize the exposure of EDC to air; secure closures on stock

and dry aliquot containers, store desiccated at -20°C. Use
aliquots immediately and discard containers after use. Make a
fresh 10 mg/mL EDC solution before each addition.

5. For best results, use microcentrifuge tubes from USA Scientific,

Inc., catalog number 1415-2500. To order, call 1 800 LAB-TIPS.