beautypg.com

Luminex 100 IS Developer Workbench Guide Version 2.3 User Manual

Page 20

background image

Luminex

100 IS Developer Workbench Guide Version 2.3

x

MAP

Technology

14

PN 89-00002-00-084 Rev. B

each test is selected when you create the template using this
product).

6. Enter the Test Name (analyte) and Test Units used for each test.

Apply Suffix Button—select to automatically apply the number
one to the existing test name. For each successive test name it
applies the test name and number at one increment higher than
the previous test name. For example: Enter a Test Name of
Protein and click Apply Suffix. The system changes the first
entry of Protein to Protein 1. It then adds the next test Protein 2,
and so on up to the number of tests specified. Repeat clicking of
the button repeats the process. For example, Protein 1 becomes
Protein 11, Protein 2 becomes Protein 12 and so on.

Apply Unit Button—select to apply the same unit value entered
on the first line to all other tests. For example, if mg/mL is
entered for the first Test Unit, then select Apply Unit to make all
tests mg/mL.

7. Enter the Number of Standards (assay) for this product and

press Enter on the keyboard.

8. Enter the Reagent Name for each standard in this product. Each

standard reagent must test for every analyte (bead set) used in the
assay.

9. Enter the Number of Controls (assay) for this product and press

Enter on the keyboard.

10. Enter the Reagent Name for each control in this product. Each

control reagent must test for every analyte (bead set) used in the
assay.

11. After you are satisfied with your entries, click Save.

12. Click Close.