“Quali” value for each test. Control products are optional, but are
recommended for judging the acceptability of batch results.

‹

To create a template for qualitative tests using standards and
controls:

1. Click Create Template on the Developer Workbench toolbar.

The Template Setup wizard opens. See Figure 10.

Figure 10. Template Setup Wizard—Enter Specific Information

2. Type the correct information in the Developing Company,

Created By, Template Name, Version No, and Description
boxes. The version number of the template is automatically
included in the template name that the lab personnel see when
they select the template to use with a kit.

3. Under Template Type, select Qualitative.

4. Enter the Sample Vol (µL) for this template. This is the amount

of sample for testing. Make the volume in the wells at least 25
µL greater than the sample volume.

5. Enter a Sample Timeout (sec) for this template to set the

maximum time that the analyzer will spend acquiring data for
each sample.

6. Enter the target XY Heater Temp (°C) for this template. To

enable this value, enter an XY Htr On or an XY Htr On (wait)
command into the template command list. See step 16.