Luminex 100 IS Developer Workbench Guide Version 2.3 User Manual
Page 31

x
MAP
Technology
Create Template
PN 89-00002-00-084 Rev. B
25
Figure 13. Select Bead IDs
13. Under Min Beads, enter the number of beads to collect for each
test. Click Apply Min Beads to apply the same minimum bead
requirement shown on the first line to all other tests.
14. Configure formulas. You can select one of the pre-defined
formulas from the Qualitative Analysis Method list or edit one of
the formulas and add it to the list under a new name. See
Figure 14.
To select an existing formula: click in the Formula column and
select a formula from the list box. Then, continue with step 16.
To add a new formula: continue with the following Qualitative
Formula Synopsis.