X-y format – B&K Precision 2552-2559 - Manual User Manual
Page 79

79
Setup Grid
Press the “Grid” option button to select “
”, “
”or“
”to set the screen whether to
display grid or not.
Setup Menu Display
Press the “Menu Display” option button to select “2
sec”, “5 sec”, “10 sec”, “20 sec”, or “Infinite” to set
menu display time on screen.
Set Skin
Press the “skin” option button or turn the “Universal”
knob to select “Classical”, “Modern”, “Traditional”, or
“Succinct”.
X-Y Format
Use the XY format to analyze phase differences, such as those represented
by Lissajous patterns. The format plots the voltage on channel 1 against the
voltage on channel 2, where channel 1 is the horizontal axis and channel 2
is the vertical axis. The oscilloscope uses the untriggered Sample
acquisition mode and displays data as dots.
Operation steps:
●
Channel 1 (or channel 3) “Volts/div” and vertical
“POSITION” set up the horizontal scale and position.
●
Channel 2 (or channel 4) “Volts/div” and vertical
“POSITION” set up the horizontal scale and position.
●
Turn the “Time/div” knob to adjust the sampling rate
●
The following functions are forbidden at XY display
form:
● Benchmark waveform and wave mathematics
●
Cursor
Note: The oscilloscope can capture a waveform in normal YT mode
at any sampling rate. You can view the same waveform in XY mode.
To do so, stop the acquisition and change the display format to XY.