Chapter 6 Specifications
•
All specifications apply to measurements with a 10X probe. To
verify that the oscilloscope meets specifications, the oscilloscope
must first meet the following conditions:
o The oscilloscope must have been operating continuously
for thirty minutes within the specified operating
temperature.
o You must perform the Do Self Cal operation, accessible
through the Utility menu, if the operating temperature
changes by more than 5° C.
o The oscilloscope must be within the factory calibration
interval.
•
All specifications are guaranteed unless noted “typical.”
Specifications
2552/2553 2554/2555
2556/2557
2558/2559
Inputs
Input
Coupling
AC, DC, GND
Input
Impedance
1 MΩ ± 2% || 13 pf ± 3 pf 1 MΩ ± 2% || 13 pf ± 3 pf
50 Ω ± 2%
Maximum
input
voltage
400 V (DC+AC PK-PK, 1 MΩ input impedance, X10),
CAT I
Ch to Ch
Isolation
(Both
channels in
same V/div
setting)
>100:1 at
35 MHz
>100:1 at
50 MHz
>100:1 at
100 MHz
>100:1 at 150
MHz
Probe
attenuator
1X, 10X
Probe
attenuations
selectable
1X, 5X, 10X, 50X, 100X, 500X, 1000X