Vspp, Vspp vertical sync pulse polarity – Quantum Data 802BT User Manual
Page 576

566
Appendix A Command Reference
VSPP
Vertical Sync Pulse Polarity
Class
Format parameter setting - Synchronization
Description
Establishes the logic sense of the digital vertical sync outputs. Setting polarity to 1 causes
the leading edge of vertical sync to be a low-to-high transition. Setting polarity to 0 causes
the leading edge of vertical sync to be a high-to-low transition. The VSPP? query returns
the current polarity of VSPP.
Command syntax
VSPP polarity
polarity
0 = high-to-low transition
1 = low-to-high transition
Query syntax
VSPP?
Query response
0 or 1
Other required
commands
In order to use digital vertical sync, it must be gated on with the
command and
digital separate H and V sync must be selected with the
command. The
command instructs the generator to use the new setting. The
command updates
hardware to the new setting and redraws the test image.
Example
VSPP 1 // Set active hi V sync in buffer
VSPG 1 // Enable V sync output in buffer
SSST 1 // Select H&V sync type in buffer
FMTU
// Update hardware to current buffer contents