Cspp, Cspp composite sync pulse polarity – Quantum Data 802BT User Manual
Page 270

260
Appendix A Command Reference
CSPP
Composite Sync Pulse Polarity
Class
Format parameter setting - Synchronization
Description
Establishes the logic sense of the digital composite sync output. The CSPP? query returns
the current setting of CSPP.
Command syntax
CSPP polarity
polarity
0 = active-low (negative going pulse)
1 = active-high (positive going pulse)
Query syntax
CSPP?
Query response
0 or 1
Other required
commands
In order to use digital composite sync, it must be gated on with the
command and
selected with the
command. The
command instructs the generator to use the
new setting. The
command updates hardware to the new setting and redraws the
test image.
Example
CSPG 1 // Enable dig comp sync in buffer
CSPP 1 // Select active hi in buffer
SSST 2 // Choose digital comp sync type in buffer
FMTU
// Update hardware to current buffer contents