Hrat, Hrat horizontal rate – Quantum Data 802BT User Manual
Page 377

Video Test Generator User Guide
367
HRAT
Horizontal RATe
Class
Format parameter setting - Video resolution
Description
Sets the line frequency. Pixel rate is equal to HTOT multiplied by HRAT. Frame rate is
equal to HRAT divided by VTOT. Field rate is equal to SCAN multiplied by the frame rate.
The HRAT? query returns the current horizontal frequency setting.
Command syntax
HRAT frequency
frequency
typical min = 1000 Hz (floating point accepted)
typical max = 130000 Hz (floating point accepted)
Query syntax
HRAT?
Query response
frequency in Hz (floating point returned)
Other required
commands
The
command instructs the generator to use the new settings. The
command
updates the hardware to the new settings and redraws the test image.
Example
HRAT 32768 // Set 32.768 kHz rate in buffer
FMTU
// Update hardware to current buffer contents
or
HRAT 32.768E3 // Set a 32.768 kHz rate
FMTU
// Update hardware to current buffer contents