Avps, Vel is set with the, Avps analog video pedestal swing – Quantum Data 802BT User Manual
Page 252

242
Appendix A Command Reference
AVPS
Analog Video Pedestal Swing
Class
Format parameter setting - Analog video signal
Description
Sets a black pedestal level between the blanking level (0.0 I.R.E.) and the peak video
level (100.0 I.R.E.). The AVPS? query returns the current setting of AVPS.
Command syntax
AVPS level
level
min = 0.0 I.R.E.
max = 100.0 I.R.E.
Query syntax
AVPS?
Query response
level
Other required
commands
must be set to ON to enable the use of the pedestal. The
command instructs
the generator to use the new setting. The
command updates hardware to the new
setting and redraws the test image.
Example
AVPG 1
// Enable use of a black level pedestal
AVPS 7.5
// Set pedestal level to 7.5 IRE
FMTU