4 calibration exit (vxi plug&play demo), 5 changing the calibration password, Table 4-1. factory default unit passwords – KEPCO KLP Series User Manual, Rev 4 User Manual
Page 90: Calibration exit (vxi plug&play demo) -10, Changing the calibration password -10, Factory default unit passwords -10, R. 4.5, E 4- 1, R 4.4.4), R. 4.4.4
4-10
KLP 091313
4.4.4
CALIBRATION EXIT (VXI plug&play DEMO)
NOTE: Clicking the Calibrate Close button on the VXI main panel will exit calibration WITH-
OUT saving any calibration data.
1. Click the SAVE CALIBRATION button on the calibration window to save this calibration.
2. The Calibration window automatically closes and the Virtual Model window automatically
opens.
3. The virtual model settings can be changed after the password has been properly entered.
4. NOTE: The virtual model is always reset following calibration. If voltage calibration is com-
pleted (regardless of whether or not current calibration is completed), the virtual model volt-
age is set for maximum rated voltage, and current is set to the value corresponding to 1200
Watts. If a current calibration is completed, but no Voltage calibration, the virtual model cur-
rent will be set for maximum rated current and voltage is set to the value corresponding to
1200 Watts.
4.5
CHANGING THE CALIBRATION PASSWORD
The 4-digit Unit password is required for to enter calibration mode. The factory default Unit
password is listed in Table 4-1. To change the password from the front panel proceed as follows:
1. If the status display shows SET, tap either the CURRENT or VOLTAGE controls to take the
unit out of setpoint mode (status display goes from SET to blank)
2. Using two thin tools, while holding the PROTECT switch in, tap the FUNCTION switch once.
Status display reads PASS.
NOTE: The FUNCTION switch must be actuated within two seconds, i.e., before PROT
appears in the status display. To exit this procedure without changing the password,
rotate the VOLTAGE control.
3. Tap the DC OUTPUT switch. Status display reads OLD?.
4. Rotate the CURRENT control until DC AMPERES display shows old password. Tap the DC
OUTPUT switch to enter. Status display shows NEW?.
TABLE 4-1.FACTORY DEFAULT UNIT PASSWORDS
MODEL
(STANDARD OR E-SERIES)
UNIT
PASSWORD
KLP 10-150
1015
KLP 20-120
2012
KLP 36-60
3660
KLP 75-33
7533
KLP 150-16
1516
KLP 300-8
3008
KLP 600-4
6004