4-10

KLP 091313

4.4.4

CALIBRATION EXIT (VXI plug&play DEMO)

NOTE: Clicking the Calibrate Close button on the VXI main panel will exit calibration WITH-

OUT saving any calibration data.

1. Click the SAVE CALIBRATION button on the calibration window to save this calibration.

2. The Calibration window automatically closes and the Virtual Model window automatically

opens.

3. The virtual model settings can be changed after the password has been properly entered.

4. NOTE: The virtual model is always reset following calibration. If voltage calibration is com-

pleted (regardless of whether or not current calibration is completed), the virtual model volt-
age is set for maximum rated voltage, and current is set to the value corresponding to 1200
Watts. If a current calibration is completed, but no Voltage calibration, the virtual model cur-
rent will be set for maximum rated current and voltage is set to the value corresponding to
1200 Watts.

4.5

CHANGING THE CALIBRATION PASSWORD

The 4-digit Unit password is required for to enter calibration mode. The factory default Unit
password is listed in Table 4-1. To change the password from the front panel proceed as follows:

1. If the status display shows SET, tap either the CURRENT or VOLTAGE controls to take the

unit out of setpoint mode (status display goes from SET to blank)

2. Using two thin tools, while holding the PROTECT switch in, tap the FUNCTION switch once.

Status display reads PASS.

NOTE: The FUNCTION switch must be actuated within two seconds, i.e., before PROT

appears in the status display. To exit this procedure without changing the password,
rotate the VOLTAGE control.

3. Tap the DC OUTPUT switch. Status display reads OLD?.

4. Rotate the CURRENT control until DC AMPERES display shows old password. Tap the DC

OUTPUT switch to enter. Status display shows NEW?.

TABLE 4-1.FACTORY DEFAULT UNIT PASSWORDS

MODEL

(STANDARD OR E-SERIES)

UNIT

PASSWORD

KLP 10-150

1015

KLP 20-120

2012

KLP 36-60

3660

KLP 75-33

7533

KLP 150-16

1516

KLP 300-8

3008

KLP 600-4

6004