3 options, 4 specifications, Options -2 – KEPCO KLP Series User Manual, Rev 4 User Manual
Page 14: Specifications -2
1-2
KLP 091313
1.3
OPTIONS
Options are identified by a letter suffix. Unless otherwise noted, options apply equally to stan-
dard and E-Series models, and specifications are not affected.
1.3.1
RAPID OUTPUT DISCHARGE CIRCUIT (RODC) OPTION (R SUFFIX)
The Rapid Output Discharge Circuit (RODC) option (suffix R) is available on all KLP models.
This circuit rapidly discharges the output capacitance, thus significantly reducing response time
to reductions in output voltage.
The circuit consists of a voltage detector that compares the programmed and actual values of
output voltage. The discharge circuit is activated only when the actual voltage exceeds the pro-
grammed value.
Without the RODC circuit, discharge of the total output capacitance (internal and external) is
achieved through a combination of the external load resistance and an internal current sink. For
high load resistance or open circuit conditions at the output, response time (fall time) can vary
from milliseconds to seconds depending upon the magnitude of the high-to-low voltage transi-
tion.
With the RODC option, output fall time is reduced to approximately the same value as rise time,
even with external capacitance equal to 50% of the nominal internal output capacitance. Table
1-1 illustrates the typical performance improvement that can be expected with an open circuit
output.
The RODC circuit will support continuous repetitive discharges from rated model voltage to zero
at rate of up to 1 Hz for the 10V, 20V and 36V models, and up to 0.5 Hz for the 75V, 150V, 300V
and 600V models. Smaller discharge steps will be supported at higher continuous rates; con-
tact Kepco Applications Engineering for additional details. The RODC circuit employs protection
against both peak energy and average power overload conditions by disabling only the RODC
function until conditions recover to normal range; unit performance remains otherwise unaf-
fected. Note that RODC Option units support the TRANsient command (see KLP Developer’s
Guide for details).
1.4
SPECIFICATIONS
Table 1-2 lists the parameters applicable to individual models. Table 1-3 lists general specifica-
tions applicable to all models except where otherwise noted.
TABLE 1-1. RAPID OUTPUT DISCHARGE CIRCUIT IMPROVEMENT IN OUTPUT RESPONSE TIME
KLP Model No.
Nominal Internal Output
Capacitance
(CO)
Test Voltage
Maximum Fall Time @ 1.5 x Co
without RODC
with RODC(1)
KLP 10-150-
(2)
28600 µF
10V
600 msec
30 msec
KLP 20-120-
(2)
13000 µF
20V
675 msec
30 msec
KLP 36-60-
(2)
7280 µF
36V
1460 msec
30 msec
KLP 75-33-
(2)
5600 µF
75V
3690 msec
25 msec
KLP 150-16-
(2)
2200 µF
150V
8940 msec
30 msec
KLP 300-8-
(2)
550 µF
300V
8720 msec
26 msec
KLP 600-4-
(2)
118 µF
600V
6500 msec
33 msec
(1) Values shown assume that an external capacitance of 1/2 value of CO is connected across the output terminals.
(2) Specifications listed apply to both standard and LXI models.