5 analog input full scale calibration, 6 calibration, 3 remote mode programming using digital interfaces – KEPCO KLP Series User Manual, Rev 4 User Manual
Page 64: 1 factory default settings, 1 gpib factory defaults, Analog input full scale calibration -22, Calibration -22, Factory default settings -22, Gpib factory defaults -22, R. 3.2.17.5)
3-22
KLP091313
4. Tap the DC OUTPUT switch to accept the new settings (blinking stops) or use a thin tool
(e.g., a paper clip) to press the FUNCTION switch to cancel the operation.
5. To exit, rotate either VOLTAGE or CURRENT control until the status display reads RTN, then
tap the DC OUTPUT switch. To exit the UTIL menu rotate either VOLTAGE or CURRENT
control until the status display reads EXIT, then tap the DC OUTPUT switch.
3.2.17.5 ANALOG INPUT FULL SCALE CALIBRATION
Enter UTIL menu (see PAR. 3.2.17, steps 1 through 3) and rotate either VOLTAGE or CUR-
RENT control until status display reads ANPR.
• The DC VOLTS display shows the analog input voltage value that produces the full scale
output voltage established by the virtual model.
• The DC AMPS display shows the analog input voltage value that produces the full scale
output current established by the virtual model.
1. To change the settings tap the DC OUTPUT switch. Note that DC VOLTS and DC
AMPERES displays begin blinking.
2. Rotate VOLTAGE control to set the analog voltage value required to produce the full scale
output voltage.
3. Rotate the CURRENT control to set the analog voltage value required to produce the full
scale output current.
4. Tap the DC OUTPUT switch to accept the new settings (blinking stops) or use a thin tool
(e.g., a paper clip) to press the FUNCTION switch to cancel the operation.
5. To exit, rotate either VOLTAGE or CURRENT control until the status display reads RTN, then
tap the DC OUTPUT switch. To exit the UTIL menu rotate either VOLTAGE or CURRENT
control until the status display reads EXIT, then tap the DC OUTPUT switch.
3.2.17.6 CALIBRATION
Refer to PAR 4.3.
3.3
REMOTE MODE PROGRAMMING USING DIGITAL INTERFACES
For digital remote programming refer to the KLP Developer’s Guide.
3.3.1
FACTORY DEFAULT SETTINGS
3.3.1.1
GPIB FACTORY DEFAULTS
The factory default GPIB address is 6; refer to PAR. 3.2.12 to change the GPIB address.