beautypg.com

Figure 4-3. voltage calibration window, 2 current calibration (vxi plug&play demo), Current calibration (vxi plug&play demo) -7 – KEPCO KLP Series User Manual, Rev 4 User Manual

Page 87: Voltage calibration window -7, R 4.4.2 for, E 4-3)

background image

KLP 091313

4-7

FIGURE 4-3. VOLTAGE CALIBRATION WINDOW

2. Click the Press to ACCEPT button to accept the value. The status display then reads

VMAX. Monitor

DVM and click the “+” button to increase and the “–“ to decrease the output

voltage until the DVM reads as close as possible to maximum rated voltage.

3. Click the Press to ACCEPT button to accept the value; the status display now reads OUTV.

At the ANALOG I/O PORT J2, connect the DVM to VOLT_RBACK and GND as shown in
Figure 4-1. Click the “+” button to increase and the “–“ button to decrease the output voltage
until the DVM reads as close as possible to 10V.

4. Click the Press to ACCEPT button to accept the value. The status display will not change.

At the calibration window, click the CURRENT Button to proceed to current calibration (see
PAR. 4.4.2)

.

4.4.2

CURRENT CALIBRATION (VXI plug&play DEMO)

During current calibration, the current, current readback and analog current readback are cali-
brated.

1. Click the CURRENT button on the Calibration Window to start current calibration; a pop-up

box appears instructing the operator to connect the shunt across the output terminals. Once
the shunt is connected, click OK; the Current Calibration Window (Figure 4-4) appears and
the status display on the VXI main panel reads CMAX.

This manual is related to the following products: