beautypg.com

Figure 4-5. analog reference calibration window, Analog reference calibration window -9, R. 4.4.3 – KEPCO KLP Series User Manual, Rev 4 User Manual

Page 89

background image

KLP 091313

4-9

6. Click the PRESS TO ACCEPT button to accept the value. The status display will not change.

Click the ANALOG REF button on the calibration window to proceed to external calibration
(see PAR. 4.4.3).

4.4.3

ANALOG REFERENCE CALIBRATION (VXI plug&play DEMO)

During analog reference calibration the external analog reference voltages used to establish full
scale output current and voltage are calibrated.

1. Click the ANALOG REF button on the Calibration Window; the Analog Reference Calibration

Window (Figure 4-5) appears and the status display on the VXI main panel reads INPV. With
the reference voltage adjusted to 10.000V, click the PRESS TO ACCEPT button to accept
the value. When the calibration is saved, this analog voltage will be equivalent to full scale
output voltage.

FIGURE 4-5. ANALOG REFERENCE CALIBRATION WINDOW

2. The status display reads INPC. With the reference voltage adjusted to 10.000V, click the

PRESS TO ACCEPT button to accept the value.

3. The status display will not change. Proceed to PAR. 4.4.4 to exit calibration.

This manual is related to the following products: