R 3.2.15 to, Ar 3.2.15 – KEPCO KLP Series User Manual, Rev 4 User Manual
Page 56
3-14
KLP091313
• Calibration cannot be performed on a master/slave pair; the individual units must be
reconfigured back to independent mode (PAR 3.2.14.4) and then calibrated as individual
units.
• Baud rate is automatically adjusted to highest supportable rate as part of master/slave
setup and cannot be adjusted by user.
• LIST, VOLT:MODE TRANS and CURR:MODE TRANS commands are not supported for
master/slave pairs.
3.2.14.4 RESTORING THE MASTER/SLAVE CONFIGURATION TO STANDALONE OPERATION
[STANDARD MODELS ONLY]
First configure the master, then the slave for standalone operation.
1. If the units are in setpoint mode (see PAR. 3.2.7), tap either the CURRENT or VOLTAGE
controls to take the units out of setpoint mode (status display goes from SET to blank).
2. Using a thin tool (e.g., a paper clip), press the FUNCTION switch of the unit to be restored to
standalone operation, then rotate either the CURRENT or VOLTAGE control until the display
reads M/S.
3. Tap the DC OUTPUT switch once.
• The status display reads either SERI (series) or PARA (parallel).
• The DC VOLTS display shows the number assigned to the unit (1 = master). The DC
AMPERES display shows the maximum number of units in the configuration.
4. To change the configuration tap the DC OUTPUT switch once. The DC VOLTS and DC
AMPERES start blinking.
• Rotate the CURRENT control until DC AMPERES display shows 1.
• Rotate the VOLTAGE control until DC VOLTS display shows 1.
5. Press DC OUTPUT to accept the displayed settings. The unit briefly flashes WAIT, then
restarts with the new settings; press FUNCTION to abort without changing the settings.
NOTE: Once the master is configured for standalone operation, the slave status display will
flash E241 (hardware error).
6. Repeat steps 2 through 5 to restore the slave unit to standalone operation.
3.2.15
LAN INTERFACE CONFIGURATION [E-SERIES MODELS ONLY]
If the status display shows SET, tap either the CURRENT or VOLTAGE controls to take the unit
out of setpoint mode (status display goes from SET to blank). Using a thin tool (e.g. end of paper
clip), press FUNCTION switch once, then rotate either VOLTAGE or CURRENT control until sta-
tus display reads LAN. Press DC OUTPUT switch to enter the LAN Interface configuration
menu. Rotate either VOLTAGE or CURRENT control to see the available functions.
• LRST - LAN Reset. To restore the LAN interface configuration to the factory default, tap
DC OUTPUT switch once, then rotate either VOLTAGE or CURRENT control to show
YES in the DC VOLTS display, then tap DC OUTPUT switch again. The factory default
settings listed in PAR. 3.3.1.3 are restored upon exiting the Function menu.