RIGOL DG4000 Series User Manual
Page 90
RIGOL Chapter 3 Arbitrary Waveform Output
DG4000 Series User’s Guide
3-14
1) Press PointID to define the first point (the time is 0 by default). Press
Voltage, use the numeric keyboard to input “0” and select the unit “V”
from the pop-up menu.
2) Press PointID again, use the numeric keyboard or knob to select point 2;
then press Time and Voltage to input 4μs and 4 V respectively.
3) Refer to step 2) to input point 3 and 4 in the table.
4) After finishing editing all the points, press OK to return to the previous
menu. At this point, press Save to enter the file storage interface (refer to
Store and Recall) to save the edited arbitrary waveform.
7. View the edited waveform.
V
μs
4
2
0
-2
4
8
12
2
1
3
4
1
Note: As can be seen from the above figure, the waveform editor will always
connect from the first point to the voltage level of the last point to create a
continuous waveform.