beautypg.com

To select pm modulation, To select carrier waveform shape, To set carrier frequency – RIGOL DG4000 Series User Manual

Page 108: Pm -8

background image

RIGOL Chapter 5 Modulated Waveform Output

DG4000 Series User’s Guide

5-8

PM


Modulated waveform consists of carrier waveform and modulating waveform. For
phase modulation (PM), the phase of the carrier waveform varies with the
instantaneous voltage of the modulating waveform.

To Select PM Modulation


Press ModTypePM to enable PM function.
 When Mod is enabled, Sweep or Burst will be automatically disabled (if

currently enabled).

 After PM is enabled, the instrument will generate PM waveform with the

currently specified carrier and modulating waveforms.

To Select Carrier Waveform Shape


PM carrier waveform could be Sine, Square, Ramp or Arbitrary waveform (except
DC). The default is Sine.
 Press the front panel key Sine, Square, Ramp or ArbSelect Wform

(or User, the pre-defined waveform can not be DC) to select the desired
carrier waveform shape.

 Pulse, Noise and DC could not be used as carrier waveform.

To Set Carrier Frequency


Different carrier waveforms have different frequency ranges as shown in the table
below. For all the carrier waveforms, the default value is 1kHz.

Carrier Waveform

Frequency Range

Sine

1μHz to 160MHz

Square

1μHz to 50MHz

Ramp

1μHz to 4MHz

Arbitrary Waveform

1μHz to40 MHz