To set parameters for pulse, Pulse width/duty cycle, To set parameters for pulse -13 – RIGOL DG4000 Series User Manual
Page 69: Pulse width/duty cycle -13
Chapter 2 Basic Waveform Output
DG4000 Series User’s Guide
2-13
To Set Parameters for Pulse
To output a pulse, users need to set the “Pulse Width/Duty Cycle”, “Leading edge”,
“Trailing edge”, “Delay” and “Recover Delay” in addition to the basic parameters
(such as frequency, amplitude, DC offset voltage, start phase, high level, low level
and align phase) introduced above.
Pulse Width/Duty Cycle
Pulse Width is defined as the time from the 50% threshold of a rising edge
amplitude to the 50% threshold of the next falling edge amplitude as shown in the
figure above.
Pulse Width range is limited by the “Minimum Pulse Width” and the “Pulse
Period” (for the ranges of “Minimum Pulse Width” and “Pulse Period”, please refer
to “Signal Characteristics” in Specifications). The default value is 500μs.
Pulse Width ≥ Minimum Pulse Width
Pulse Width ≤ Pulse Period - Minimum Pulse Width × 2
Pulse Duty Cycle is defined as the percentage that the Pulse Width takes up in the
whole period.
脉宽
10%
t
Rise
t
Fall
90%
50%
Pulse Period
Pulse Width